Lang: VN
Hotline: 028.6676.7762
icon-phone icon-mail icon-zalo

REALTIME PCR

Real-time PCR LineGene Mini S, Bioer (FQD-16B EA4, FQD-16B EA2)

LineGene Mini s
Brand: Bioer
Model: FQD-16B

Price: Call

Line Gene Mini S is a portable fluorescence quantitative PCR instrument developed by the Bioer company to deal with diversified and complex environment. This product adopts the small capacity design of 16 single-tube , and can be used for battery power supply.At the same time, this new generation of PCR is used to create a new generation of fluorescence quantitative PCR with a customized version of Peltier module, new optical path design and top scanning.
Application environment:  1,field operation, such as field, forest farm, farm and water source.2, disaster area or the field hospital, to make rapid and accurate clinical diagnosis of the disaster. Of course, it is also convenient for the handling and use of the lab.

 

BYQ 6618E BYQ 6622E BYQ 6623E
Fl: FAM, SYBR Green I
F3 : ROX
Fl: FAM, SYBR Green I
F2 : VIC, HEX, TET, JOE, TAMRA, CY3, NED
Fl : FAM , SYBR Green I
F2 : VIC, HEX, TET, JOE, TAMRA, CY3, NED
F3 : ROX
F4 : CY5
Parameter LineGene-Mini
Model FQD-16A
Sample Capacity 16 x 0.2ml tubes (flat and transparent cap)
2 x 8-tube strips (flat and transparent cap)
Temperature Control Method Peltier
Temp. Control Mode Tube mode
Reaction Volume 10-100ul
Temp.Range 0 – 100°C
Maximum Heating/Cooling Rate 5°C/s
Average Heating/Cooling Rate 3.5°C/s
Temperature Resolution 0.1°C
Hot Lid Temperature Range 85-110°C
Temp. Accuracy ± 0.1°C
Temp.-Uniformity ± 0.15°
Excitation Wavelength Range 400 to 700nm
Detection Wavelength Range 450 to 750nm
Sensitivity 1 copy
Dynamic Range 1-1010 Copies
Dimension W 280 x D 220 x H 240
Net Weight 6.5 kg
Adapter Input Voltage 100 – 240V
Adapter Input Frequency 50 – 60Hz
Power DC 24V 180 W, power adaptor needed
Connection Options USB port
Safety Protection and Alarm Block, hot lid and heat sink temperature sensors short-circuit, open-circuit alarm and protection
Hot lid over temperature alarm and protection
Environment Temperature Range 5-35°C
Safety Certification IVD, CE, MET, RoHS 2.0
Warm-up Time No